Chapter 3 Assembly & Installation
INSTALLATION INSTRUCTIONS
Assembly of Models QT-710 and QT-720 are identical, the only difference is
Model QT-710 has legs with factory-installed wheels. The illustrations that fol-
low show assembly of a Model QT-720 radiographic table as an example, how-
ever, the procedure is same for both models. Because Model QT-720 requires
fastening to the floor (Model QT-710 is mobile), it is necessary to perform the
"Table Alignment" procedure at the end of this chapter to complete installation
of this model.
Model QT-710 and Model QT-720 Table Assembly
1. Attach the Table Leg Brace to rear side of each Table Leg using four (4)
1/4-20 x 3/4" hex bolts (see Figure 5).
2. Place Table Top Frame on Table Legs and secure using four (4) 5/16-18
UNC x 1" socket head cap screws. 5/16" flat washers, and 5/16" lockwash-
ers.
14
5/16-18 X 1"
SOCKET HEAD
CAP SCREWS
(4X)
Figure 5. Table Assembly (Viewed from Rear)
1/4-20 X
3/4"HEX
HEAD
BOLTS
Model QT-710/720 Radiographic Table
NOTE
Prior to install-
ing the Model
QT-720 Radio-
graphic Table,
the area of the
floor on which
the Table Legs
are to be
mounted must
be level in all
directions.
Need help?
Do you have a question about the QT-710 and is the answer not in the manual?