Download Print this page

Quantum QT-710 Installation And Operation Manual

Mobile/fixed non-float-top open-base radiographic table

Advertisement

Models QT-710 & QT-720 Series
Mobile/Fixed Non-Float-Top
Open-Base Radiographic Table
Installation and Operation
Manual
Manual Part No. DC30-004 Revision B

Advertisement

loading

Summary of Contents for Quantum QT-710

  • Page 1 Models QT-710 & QT-720 Series Mobile/Fixed Non-Float-Top Open-Base Radiographic Table Installation and Operation Manual Manual Part No. DC30-004 Revision B...
  • Page 2 This manual is copyrighted and all rights are reserved. No portion of this document may be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine readable form without prior consent in writing from Quantum Medical Imaging, LLC.
  • Page 3 Revision History REVISION HISTORY REVISION DATE TYPE OF MODIFICATION 9/1/00 Initial Release. 2/19/01 Added UL Mark LIST OF EFFECTIVE PAGES Page Page Page Number Number Number i - x 1 - 34 Model QT-710/720 Radiographic Table...
  • Page 4 Revision History THIS PAGE INTENTIONALLY LEFT BLANK Model QT-710/720 Radiographic Table...
  • Page 5: Table Of Contents

    Table of Contents Table of Contents Section 1, Introduction Section 2, Specifications Section 3, Assembly & Installation Section 4, Operation Section 5, Maintenance Model QT-710/720 Radiographic Table...
  • Page 6 Table of Contents THIS PAGE INTENTIONALLY LEFT BLANK Model QT-710/720 Radiographic Table...
  • Page 7 Safety Notices WARNING Quantum Medical Imaging, LLC disclaims all responsibility from any in- jury resulting from improper application of this equipment. This equipment is sold to be used exclusively under the prescribed direc- tion of a person who is licensed by law to operate equipment of this na- ture.
  • Page 8 Safety Notices REGULATORY COMPLIANCE This certified Quantum Medical Imaging, LLC medical device has been designed, manufactured, and calibrated to comply with governing Federal Regulations 21 CFR Subchapter J and the performance standards attendant thereto. Upon installation, all certified products require the filing of Form FD-2579 "Report of Assembly of a Diagnostic X-Ray System"...
  • Page 9 Warranty Information WARRANTY STATEMENT Quantum Medical Imaging, LLC (herein known as “QMI”) warrants to buyer that any new product manufactured by QMI will be free from defects in material and manufactur- ing and conform substantially to applicable specifications in effect on the date of ship- ment when subjected to normal, proper and intended usage by properly trained person- nel.
  • Page 10 2) Failure of Buyer/Dealer to provide the proper incoming power required to sup- port the equipment in accordance with the recommendation of QMI. 3) Any modification of product performed by a party other than Quantum Medical Imaging. 4) Combining incompatible products.
  • Page 11 B) By making available any necessary repaired or replacement parts or as- semblies. Quantum Medical Imaging shall have the option to furnish either new or ex- change replacement parts or assemblies. All returned parts shall become the property of Quantum Medical Imaging.
  • Page 12 Warranty Information THIS PAGE INTENTIONALLY LEFT BLANK Model QT-710/720 Radiographic Table...
  • Page 13: Section 1, Introduction

    Chapter INTRODUCTION...
  • Page 15 (with or without ACL) 350 lbs. maximum patient load rating • INTENDED USE Model QT-710 and QT-720 Radiographic Tables are intended for use as a pa- tient support device during the performance of radiographic examinations. NOTE MAIN COMPONENTS Most tables See Figures 1 and 2.
  • Page 16 Chapter 1 Introduction Figure 1. Mobile Non-Float-Top Radiographic Table, Model QT-710 Series Figure 2. Fixed Non-Float-Top Radiographic Table, Model QT-720 Series Model QT-710/720 Radiographic Table...
  • Page 17: Section 2, Specifications

    Chapter SPECIFICATIONS...
  • Page 19 14 x 17 fixed grid Model QT-720: 14 x 17 fixed grid 14 x 17 bucky (with or without ACL) Bucky Weight (w/grid • and cassette tray): 35 - 40 lbs. (typical) Collimator compatibility: Manual • Automatic (Model QT-720 only) Model QT-710/720 Radiographic Table...
  • Page 20 Chapter 2 Specifications Figure 3. Model QT-710 Dimensions PERFORMANCE SPECIFICATIONS The following are performance specifications for the Radiographic Table: Image Receptor Travel Specifications • Longitudinal Travel: 40.75 inches (1035.05 mm) System Operating Environment Ambient Temperature: +10°C to +40°C • Relative Humidity: 20 to 80%, non-condensing •...
  • Page 21 • to sea level COMPATIBILITY STATEMENT The Quantum Medical Imaging, LLC Radiographic Table is compatible with all Quantum Medical Imaging, LLC manufactured tubestands, wall stands, and high-voltage x-ray generators, and with other manufacturer’s equipment with equivalent means for indication of SID and perpendicularity.
  • Page 22 Chapter 2 Specifications THIS PAGE INTENTIONALLY LEFT BLANK Model QT-710/720 Radiographic Table...
  • Page 23: Section 3, Assembly & Installation

    Chapter ASSEMBLY & INSTALLATION...
  • Page 25 The Radiographic Table is shipped in separate shipping containers as follows: transporting agent and ask for an Table Legs (with attached wheels on Model QT-710), Table Top Assembly, • "Inspection of Table Leg Brace, and, if ordered, a Receptor Cabinet (i.e., fixed grid cabinet Damage".
  • Page 26 Because Model QT-720 requires Prior to install- fastening to the floor (Model QT-710 is mobile), it is necessary to perform the ing the Model “Table Alignment” procedure at the end of this chapter to complete installation QT-720 Radio- of this model.
  • Page 27 Through the opening in the top of the table, insert the Front Receptor Bearings so they rest on the front bearing track (see Figure 7). Care- fully, lower the back end of the Receptor Cabinet so it rests on the Ta- ble Leg Brace (see Figure 8). Model QT-710/720 Radiographic Table...
  • Page 28 Chapter 3 Assembly & Installation FRONT RECEPTOR BEARING BRACKET WHEELS ON RAIL Figure 7. Front Receptor Bearings Resting on Rail TABLE LEG BRACE Figure 8. Receptor Cabinet Resting on Table Leg Brace Model QT-710/720 Radiographic Table...
  • Page 29 TRACK THROUGH SLOT IN BEARING RAIL Figure 9. Insert Left and Right Receptor Bearings on Track LEFT AND RIGHT REAR RECEPTOR BEARING BRACKETS TABLE LEG BRACE Figure 10. Position Receptor Cabinet for Rear Receptor Bracket Installation Model QT-710/720 Radiographic Table...
  • Page 30 Provide a service loop in cable such that there is no cable interference at any point throughout the full travel of Receptor Cabinet. Connect system ground wire to the ground stud located on rear side of left table leg. Model QT-710/720 Radiographic Table...
  • Page 31 If the tubestand to be used provides transverse travel of the x-ray tube relative to the table, set the tube transverse travel in the center detent position prior to performing this alignment. 1. Position the table under the collimator, in the approximate location where it will be mounted. Model QT-710/720 Radiographic Table...
  • Page 32 All mounting procedures must be approved before, and checked after, installa- tion by the hospital or facility engineering department and a registered, profes- sional, structural engineer. It is recommended that a leakage current measure- ment be taken following installation to verify system meets allowable levels. Model QT-710/720 Radiographic Table...
  • Page 33: Section 4, Operation

    Chapter OPERATION...
  • Page 35 TABLE WHEEL LOCK OPERATION (MODEL QT-710 ONLY) On Model QT-710 mobile radiographic tables, wheel locks are located on the front left and right wheels of the table for immobilizing the table (see Figure 12). To operate the wheel locks, proceed as follows: Locking the Wheel - Press down on the end of the lever on a locking wheel.
  • Page 36 To load a film cassette into a “Midwest” type cassette tray, proceed as fol- lows: 1. Pull cassette tray from Receptor Cabinet using tray handle (see Figure 14). 2. Pull back on front cassette grip (opens both grips). Model QT-710/720 Radiographic Table...
  • Page 37 5. Lock cassette into tray. 6. Push front grip against cassette. 7. While pushing front grip against cassette, turn cassette lock handle to lock position. 8. Push tray into Receptor Cabinet. The cassette is now in exposure posi- tion. Model QT-710/720 Radiographic Table...
  • Page 38 3. Grasp handle and slide clamp back for inserting cassette. 4. Insert cassette into tray (back end first). Center it using centering la- bel. 5. Slide clamp forward pushing clamp firmly against cassette. 6. While pushing clamp against cassette, press clamp handle down Model QT-710/720 Radiographic Table...
  • Page 39: Section 5, Maintenance

    Chapter MAINTENANCE...
  • Page 41 Verify that the equipment is properly grounded before a area. Verify that the equipment is properly grounded before a area. Verify that the equipment is properly grounded before tempting any electrical operation. tempting any electrical operation. tempting any electrical operation. Model QT-710/720 Radiographic Table...
  • Page 42 2. Rotate Delrin Brake (eccentric) to either increase or decrease gap between it and table frame. 3. Re-tighten socket head cap screw. 4. Re-check operation of Receptor Cabinet Lock Release Lever over entire travel of Receptor Cabinet. Repeat adjustment if necessary. Model QT-710/720 Radiographic Table...
  • Page 43 Table Wheel, 5” Locking (Model QT- ME41-001 710 only) Table wheel, 5” Non-Locking (Model ME41-002 QT-710 only) Wheel Mounting Hardware (2-13 x 1” FHN-1316-3 hex head screw) (Model QT-710 only) Wheel Mounting Hardware (1/2-13 FFW-12N-3 lockwasher) (Model QT-710 only) Model QT-710/720 Radiographic Table...
  • Page 44 Chapter 5 Maintenance Figure 17. Replaceable Parts Location Diagram Figure 18. Replaceable Parts Location Diagram Model QT-710/720 Radiographic Table...
  • Page 45 Chapter 5 Maintenance BEARING BRACKET (TYPICAL) Figure 19. Replaceable Parts Location Diagram Figure 20. Replaceable Parts Location Diagram Model QT-710/720 Radiographic Table...
  • Page 46 Quantity required • When ordering components or parts not listed in Table 1, a complete description of the part, including its function and location should be provided with the model number and serial number of the unit. Model QT-710/720 Radiographic Table...