Change the output unit from bar to inches of water column
Change the output unit from °Celsius to °Fahrenheit
Change the output from level above sensor to depth to water surface
The standard setting for the Chatter™ Data Logger is to
measure the water level above the pressure transmitter, A in
the drawing to the right
If you need to get the depth to water instead, B in the
drawing to the right, you will have to change the settings.
You must know the sum of A and B, it equals the depth of the
sensor tip measured from the soil surface.
In the examples below, A is set to 3 meters and B is set to 5
meters, the sum is 8 meters.
The depth to water surface is always a negative
number!
Output value unit is in bar
Output value (401,4742 1 Sensor value) (0 0)
Output value Sensor value 401,4742
Output value (1 1,8 Sensor value) (32 0)
Output value Sensor value 1,8 32
176
USERS MANUAL
CHATTER™ DATA LOGGER