Outline Of Control Commands - KYOWA WGA-680A Instruction Manual

Instrumentation amplifier (for bcd/rs-232c/rs-485/teds options)
Hide thumbs Also See for WGA-680A:
Table of Contents

Advertisement

2-5. OUTLINE OF CONTROL COMMANDS

This section describes the format, type, etc. of the control commands, output from the external devices (PC, etc.) to the
products.
To set control commands of the products, use external devices.
(1) Control command codes
● The control commands are ASCII codes and 3 characters.
● The control commands may include parameters based on the command characteristics.
● The control commands include the termination code as the exit code of the control commands.
The termination code varies with the operations.
<CR>
<CR><LF>
● When entering undefined commands, numerical values, and different decimal point position from the calibration
value, the product replies the character string in the following format.
W□□□□□□?<CR><LF>
● When an error occurs while writing the pre-set values in the EEPROM, the product replies the following format.
ng<CR><LF>
(2) Delimiters
● The delimiter appears as the exit code of the control commands.
● The delimiter varies with the operations.
<CR>
<CR><LF>
Outputs commands to products.
Outputs data from the products.
Outputs commands, from the external devices to products.
Outputs data, from the products to external devices.
18
IM-A-1060G

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents