KYOWA WGA-680A Instruction Manual page 27

Instrumentation amplifier (for bcd/rs-232c/rs-485/teds options)
Hide thumbs Also See for WGA-680A:
Table of Contents

Advertisement

(19) [EID]
Contents:
Format:
Parameter:
Return value format:
Description and others:
Reading command:
Return value:
(20) [SSM]
Contents:
Format:
Parameter:
Return value format:
Description and others:
Reading command:
Return value:
To execute TEDS sensitivity registering calibrations
EID<CR>
None
rd-*<CR><LF>
(*: Indicates the number of transducers which succeeded in connecting.)
E-03<CR><LF> (When an error occurs)
Executes the sensitivity registrations and sets the bridge excitation voltage by using
some TEDS data (rated capacity, rated output, and recommended bridge excitation
voltage).
PRR010
±x.xxxx, ±yyyyyy<CR><LF>
Rated value: -99999 to 99999 (count)
Rated output: -3.2000 to 3.2000 (mV/V)
To set smoothing functions
SSMaa, b, c<CR>
aa, b, c
Moving average
00: None
01: 2 times
02: 4 times
03: 8 times
04: 16 times
05: 32 times
06: 64 times
07: 128 times
08: 256 times
09: 512 times
10: 1024 times
11: 2048 times
ok<CR><LF>
Corresponding to the SAVE command [/M].
PRR020
aa, b, c<CR><LF>
Minimum scale
The number of moving averaging times
The value, including the decimal point. When the value has
no decimal point, it is filled with "0s" from the head.
Minimum scale
0: 1
1: 2
2: 5
3: 10
4: 20
5: 50
6: 100
7: 200
8: 500
9: 1000
Analog filter
26
IM-A-1060G
Analog filter
0: 1 Hz
1: 30 Hz
2: 100 Hz
3: 300 Hz

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents