(28) [SCV]
Contents:
Format:
Parameter:
Return value format:
Description and others:
Reading command:
Return value:
Reading command:
Return value:
Reading command:
Return value:
Reading command:
Return value:
(29) [SCV]
Contents:
Format:
Parameter:
Return value format:
Description and others:
Reading command:
Return value:
To set comparator values
SCVa, ±xxxxxx<CR>
a, ±xxxxxx
Comparator value: -99999 to 99999 (count)
The value includes the decimal point. When the value has no
decimal point, it is filled with "0s" from the head.
Comparator
1: [HI] comparator (comparator1)
2: [LO] comparator (comparator2)
3: [HH] comparator (comparator3)
4: [LL] comparator (comparator4)
ok<CR><LF>
Corresponding to the SAVE command [/M].
PRR051
[HI] comparator (comparator-1)
±xxxxxx<CR><LF>
The value, including the decimal point.
When the value has no decimal point, it is filled with "0s" from the head.
PRR052
[LO] comparator (comparator-2)
±xxxxxx<CR><LF>
The value, including the decimal point.
When the value has no decimal point, it is filled with "0s" from the head.
PRR053
[HH] comparator (comparator-3)
±xxxxxx<CR><LF>
The value, including the decimal point.
When the value has no decimal point, it is filled with "0s" from the head.
PRR054
[LL] comparator (comparator-4)
±xxxxxx<CR><LF>
The value, including the decimal point.
When the value has no decimal point, it is filled with "0s" from the head.
To set hysteresis
SCV0, xxxxxx<CR>
xxxxxx
Hysteresis: 0 to 99999
The value includes the decimal point. When the value has no
decimal point, it is filled with "0s" from the head.
ok<CR><LF>
Corresponding to the SAVE command [/M].
PRR050
xxxxxx<CR><LF>
The value, including the decimal point. When the value has no
decimal point, it is filled with "0s" from the head.
30
IM-A-1060G