KYOWA WGA-680A Instruction Manual page 30

Instrumentation amplifier (for bcd/rs-232c/rs-485/teds options)
Hide thumbs Also See for WGA-680A:
Table of Contents

Advertisement

(26) [OP1]
Contents:
Format:
Parameter:
Return value format:
Description and others:
Reading command:
Return value:
(27) [SMF]
Contents:
Format:
Parameter:
Return value format:
Description and others:
Reading command:
Return value:
To set analog outputs (D/A)
OP1020, a, ±xxxxxx, ±yyyyyy<CR>
a, ±xxxxxx, ±yyyyyy
ZERO indicated value: ±99999 (count)
Automatic scaling
0: Manual setting
1: Automatic setting
ok<CR><LF>
Corresponding to the SAVE command [/M].
PRR041
a, ±xxxxxx, ±yyyyyy<CR><LF>
ZERO indicated value
Automatic scaling
To set Hold modes
SMFa, b, c, x.xx, y.yy<CR>
a, b, c, x.xx, y.yy
Display mode
0: Normal display mode
1: Hold display mode
Hold mode
0: tracking
1: Arbitrary point hold
2: Peak Hold
3: Block-specified peak hold
4: Time-specified peak hold
5: Bottom hold
6: Block-specified bottom hold
7: Time-specified bottom hold
8: Peak-to-peak hold
9: Block-specified peak-to-peak hold
A: Time-specified peak-to-peak hold
ok<CR><LF>
Corresponding to the SAVE command [/M].
PRR070
a, b, c, x.xx, y.yy<CR><LF>
Display mode
Hold mode
FULL SCALE indicated value: ±99999 (count)
The value includes the decimal point. When the value
has no decimal point, it is filled with "0s" from the head.
FULL SCALE indicated value
Delay time: 0.00 to 9.99 (sec)
Comparison mode
Detect time
Delay time
Comparison mode
29
IM-A-1060G
Detect time: 0.01 to 9.99 (sec)
0: Normal comparison mode
1: Hold comparison mode