7
Maintenance
Introduction to Maintenance and Repair

Exchanging the Fuses of the Power Inline Filter

When
When defective
Tools required
None
Parts required
#
2
Use of unsupplied cables
WA R N I N G
Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
The Agilent 1260 Infinity micro vacuum degasser power supply has wide-ranging capability
N O T E
(see
Consequently there is no voltage selector in the rear of the Agilent 1260 Infinity micro
vacuum degasser. There are two externally accessible fuses, that protect the power supply.
These fuses are identical for all accepted line voltages.
1 Switch off the power switch at the front of the instrument.
2 Remove the power cable from the power connector at the rear of the
60
p/n
2110-0458
Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
Table 1
on page 11). It accepts any line voltage in the range mentioned in the table.
instrument.
Description
Fuse: 250V, T 500 mA , compatible to all supported line voltages