Bronchial Suctioning - Dräger Evita XL Pocket Manual

Hide thumbs Also See for Evita XL:
Table of Contents

Advertisement

Operation
Overview of the availability of additional
settings
Ventilation
mode
ATC
(optional)
X
SIMV
X
X
ILV
X
Master
ILV
X
Slave
X
PCV+ Assist
X
X
X
X

Bronchial suctioning

Before suctioning
1
Touch Special Procedure ... button in the main
menu bar.
A
The Additional Function page (A) appears by default.
2
Touch O
suction button (B).
2
3
The oxygenation procedure is started.
Evita XL ventilates the patient in the set ventilation
mode with the appropriately increased O
on.
The preoxygenation phase with the remaining time in
seconds is displayed continuously in the header bar.
20
Apnea venti-
Flowtrigger
lation
X
X
X
X
X
X
X
X
X
X
X
X
X
B
concentrati-
2
Additional settings
Inspiratory
AutoFlow
termination
X
X
X
X
X
X
X
Preoxygenation lasts for a maximum of 180 seconds.
During this time, Evita XL waits for the disconnection
necessary for suctioning. If no disconnection is detected
within 180 seconds, Evita XL terminates the oxygenati-
on procedure.
During suctioning
The time available for suctioning is displayed in se-
conds in the header bar. If suctioning is completed and
the patient reconnected within the time available,
Evita XL will end the disconnection phase.
After suctioning
After reconnection, Evita XL resumes ventilation in the
set ventilation mode. For postoxygenation, the O
centration is increased for the first 120 seconds.
The postoxygenation phase with the remaining time in
seconds is displayed in the header bar.
Canceling oxygenation
suction button.
2
Sigh
PLV
X
X
X
X
X
X
X
X
con-
2
Pocket Guide Evita XL

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents