Make a New Automated Test

The Calibration System comes with a full set of automated test files to fully calibrate a
UUT to the limits in C37.118.1a - 2014. These files can be used as templates for special
user defined test sequences. Individual tests can be modified or constructed using the
Customize and Run Single Tests feature described in Chapter 6.
To make a new automated test sequence of already existing individual tests:
1. Click Test in the top Navigation Menu .
2. Click Setup Automated Test to open the Setup Automated Test window.
3. Use the buttons on the menu to set up the automated test sequence (see Table 5-1).
Item
Load
Insert Above
Move Up
5-8
Table 5-1. Setup Automated Test Window
Name
Load an automated test list.
Insert a single test into the list.
Move the selected test up one position.
Description
gtu129.emf