Transfer T O Individual; Container - Fresenius Medical Care 460018 Operator's Manual

Dry acid dissolution unit
Table of Contents

Advertisement

7.3 T
T
I
RANSFER
O
NDIVIDUAL
1. To transfer product into individual containers (containers must be properly labeled):
A. Connect the Transfer Hose to
the Transfer Nozzle.
B. Place the Transfer Nozzle into
the opening of an individual

container.

C. Slightly OPEN the Main Transfer
Ball Valve on the top of the
Filter Housing.
D. With the Transfer Light flashing,
Press Dissolution START button.
The Transfer Pump will start.
E. Slowly OPEN Ball Valve on the
Transfer Nozzle until the desired
rate of flow through the nozzle
is achieved.
2. Once the manual TRANSFER Operation has been completed
and the individual containers are filled, CLOSE Transfer Nozzle
Valve. Press the PAUSE button.
3. When the Dry Acid Dissolution Unit is empty, the Dry Acid
Dissolution Unit will step to the CYCLE COMPLETE Operation.
CLOSE the Dry Acid Dissolution Unit's Main Transfer Ball
Valve. Then Remove Transfer Nozzle and place Transfer Hose
onto the Transfer Hose Holder of the Tank. Ensure you place
the Transfer Nozzle in appropriate area free from dirt and
other contaminates.
CAUTION! I
NOTE: A
CID CONCENTRATE SHOULD NOT BE STORED IN MIX TANK LONGER THAN
A
CID CONCENTRATE MAY BE STORED FOR LONGER PERIODS IN APPROVED CONTAINERS UNDER PROPER STORAGE
.
CONDITIONS
22
7 . 3
T R A N S F E R
C
:
ONTAINER
F THE PUMP IS RUN DRY FOR LONGER THAN
T O
I N D I V I D U A L
nd
2
Transfer Nozzle
th
5
– Slowly
Open to
desired rate
st
1
-Connect
Transfer Line to
Transfer Nozzle
30
SECONDS THIS MAY CAUSE
C O N T A I N E R S
th
4
– Press
Dissolution
START button
rd
3
– Open
Main Transfer
Ball Valve
Transfer Line
P
.
UMP FAILURE
14
DAYS FROM THE DATE OF MIXING
P/N 460018 Rev. D
.