Fresenius Medical Care 460018 Operator's Manual page 30

Dry acid dissolution unit
Table of Contents

Advertisement

1. Turn the power switch ON and ensure that the water supply valve is OPEN then press the Rinse START
button. The Fill Indicator LIGHT will illuminate indicating the RINSE CYCLE has started and the Dry Acid
Dissolution Unit is in the first FILL Operation (Figure 19).
2. Once the 25-Gallon Sensor is reached, the control panel will automatically step to the first RECIRCULATE
Operation (Figure 20). With your safety glasses on, lift the Access lid and check for correct spray ball
operation. See figure 4 on page 12, for the illustration of correct or incorrect spray ball operation.
3. When the Fill Indicator Light switches to the Recirculate Indicator Light during the second RINSE
Operation, pause the RINSE CYCLE by pressing the PAUSE button (the PAUSE Light will turn ON), then add
0.5 gallons (1.9 liters) of regular bleach (sodium hypochlorite 5% to 10%) to the rinse water in the tank.
Once the 0.5 gallons of regular bleach is added, press the RINSE CYCLE START button (the PAUSE Light will
turn off) and the RINSE Operation will continue for the duration of the 2nd RINSE Operation.
4. Once the Dissolution Unit is in the RINSE CYCLE's CYCLE COMPLETE Operation mode, initiate and
complete two RINSE CYCLEs consecutively by following the steps in Section 6, page 11.
5. When the two RINSE CYCLEs are complete, the unit should be in the RINSE CYCLE, CYCLE COMPLETE
Operation. Use the appropriate chlorine test strips to check for the proper chlorine level at the two
outputs of the Dry Acid Unit (transfer hose and drain hose).
6. Checking bleach residuals from the Transfer hose.
Press DISSOLUTION CYCLE side START button.
When water has reached the 25-Gallon Sensor, press STEP MODE ON/OFF button until the STEP MODE
indicator light turns ON.
Use STEP button to step to TRANSFER Operation, the Transfer Indicator Light will be flashing.
Have chlorine test strip ready for use
Press the DISSOLUTION CYCLE side START button. The TRANSFER Indicator Light will stop flashing, the
Transfer Valve will OPEN, and the Pump will turn ON.
OPEN the Main Transfer Ball Valve on the Dry Acid Dissolution Unit.
With the end of the Transfer Nozzle placed over a drain, slowly OPEN the valve on the Transfer Nozzle.
Allow water to flow to the drain for 15 seconds.
Adjust the output of the Transfer Nozzle to a slow output flow.
Place test strip under the flow for 30 seconds or the time interval recommended by the manufacture of
the test strip being used.
Close Transfer Nozzle Valve and press the PAUSE button (the PAUSE indicator light will illuminate).
Ensure that you check the results of the test strip immediately. If the results are higher than <0.1 ppm
(RD61: 2006) go to step 9
Before proceeding to step 8, ensure the Drain Hose is secure to prevent the hose from flailing around
during the DRAIN Operation.
24
Pause
Button
Figure 19
M A I N T E N A N C E
The Dry Acid
Dissolution
Unit will fill to
the 25-Gallon
Sensor
25-
Gallon
Sensor
Figure 20
C o n t i n u e d
P/N 460018 Rev. D