7 . 2
T R A N S F E R
7.2 T
T
S
RANSFER
O
TORAGE
CAUTION! C
OMPARE THE
D
A
C
RY
CID
ASE TO BE CERTAIN THEY ARE THE SAME
After you have obtained a valid specific gravity reading, remove the Transfer Hose from the
1.
Transfer Nozzle.
Connect Transfer Hose to the storage tank and OPEN Main Transfer Ball Valve at the side of
2.
the Filter Housing. Press the Dissolution START button to transfer concentrate into the
storage tank (Figure 10).
Once Transfer of solution is complete, CLOSE the Main Transfer Valve, and place Transfer Hose
3.
on to the Dry Acid Dissolution Unit Transfer Hose Holder (see Figure 12). Do not leave Transfer
Hose attached to the Storage Tank.
CAUTION! I
F THE PUMP IS RUN DRY FOR LONGER THAN
P/N 460017 Rev. D
T O
S T O R A G E
T
:
ANK
CATALOG NUMBER
Compare Catalog
Number on
Storage Tank
to Catalog Number
on Fresenius
Medical Care Dry
Acid Product.
st
1
- Connect Transfer
Hose to storage tank
Transfer hose
T A N K
ON THE STORAGE TANK WITH THE
.
rd
3
- Press
Dissolution
Figure 10
START button
30
SECONDS IT MAY CAUSE
CATALOG NUMBER
Transfer
Hose Holder
nd
2
- OPEN
Main Transfer
Ball Valve
P
.
UMP FAILURE
ON THE
21