R I N S E
C Y C L E
4. When the RECIRCULATE Operation is complete the Dry Acid Dissolution
Unit will switch to DRAIN Operation and empty rinse water from the Dry
Acid Dissolution Unit to the floor drain. This is a ten minute timed cycle.
5. At the completion of the DRAIN Operation, the Dry Acid Dissolution Unit
will refill to the 25-Gallon Sensor. The Fill Indicator Light will turn on and
the RINSE Operation will start.
6. When the second RINSE CYCLE is finished, the Dry Acid Dissolution Unit
will go to the CYCLE COMPLETE Operation. The Rinse Cycle Complete
Indicator Light will turn ON and the Drain Valve will remain OPEN, allowing
any residual rinse water to go down the drain.
P/N 460018 Rev. D
Continued
13