portable and mobile RF communications equipment and the 132 Gallon Dry Acid Dissolution Unit.
The 132 Gallon Dry Acid Dissolution Unit is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the 132 Gallon Dry Acid Dissolution Unit can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the 132 Gallon Dry Acid Dissolution Unit as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
68
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
d = 2.3 √P
0.23
0.73
2.3
7.3
23
P/N 460018 Rev. D