MODEL 54eA
The Model 54eA analyzer leaves the factory configured for use with the Model 499ADO sensor
(ppm dissolved oxygen). If a 499ADO sensor is not being used, turn to Section 5.5 and config-
ure the analyzer for the desired measurement (ppb oxygen, oxygen measured using a steam-
sterilizable sensor, free chlorine, total chlorine, monochloramine, or ozone) before wiring the
sensor to the analyzer. Operating the analyzer and sensor for longer than five minutes while the
analyzer is improperly configured will greatly increase the stabilization time for the sensor.
Be sure to turn off power to the analyzer before wiring the sensor.

3.3 SENSOR WIRING

3.3.1 General
The wiring label, which is shown in Figure 3-2,
is a general purpose label. It has wiring infor-
mation concerning other sensors, for example,
contacting and inductive conductivity sensors,
that can be used with the 54e instrument plat-
form. For amperometric measurements, only
TB3 and TB5 are used. Wire the amperometric
sensor to TB3. Wire the pH sensor, if one is
being used, to TB5.
3.3.2 Wiring Model 499A oxygen, chlorine,
and ozone sensors
All 499A amperometric sensors (499ATrDO,
499ADO, 499ACL-01, 499ACL-02, 499ACL-03,
and 499AOZ) have identical wiring.
Use the pigtail wire and wire nuts provided with
the amperometric sensor when more than one
wire must be attached to a single terminal.
See Figures 3-3 and 3-4.
DWG. NO.
REV.
40499A23
A
FIGURE 3-3. Amperometric sensors
with standard cable.
10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
NOTE
FIGURE 3-2. Wiring Label
DWG. NO.
REV.
40499A24
A
FIGURE 3-4. Amperometric sensors
with optimum EMI/RFI cable or Variopol cable.
SECTION 3.0
WIRING
DWG. NO.
REV.
40054e03
A