Emerson Rosemount 54eA Manual page 36

Amperometric hart analyzer/controller
Table of Contents

Advertisement

MODEL 54eA
5.5 CHOOSING DISPLAY OPTIONS (CONTINUED)
Temp units: °C
Output 1: mA
Output 2: mA
Exit
Language: English
Display left: I
Display right: Out 2
Exit
Display contrast: 40
Timeout: On
Timeout value: 10 min
Exit
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
8. Set the remainder of the display parameters. Use the
to choose the desired parameter. Then press Edit (F4). Use the
key to move the cursor to the desired selection. Press Save (F4) to
store.
Temp units
Edit
Output 1
Output 2
Language
Display left
Display right
Edit
Display Contrast
Timeout
Timeout value
Edit
Polling address
The controller uses the timeout value to activate security. Once
the controller is unlocked by entering a security code, security will
not re-activate until a display timeout occurs. If timeout has been
turned off, security will never reactivate.
SOFTWARE CONFIGURATION
°C or °F
mA or % of full scale
mA or % of full scale
English, Français, Español, Deutsch, Italiano
sensor current (I), pH (appears only if free chlorine
is being measured and pH was enabled), alarm 1
setpoint (no units), alarm 3 setpoint (no units),
or blank
sensor current (I), pH (appears only if free chlorine
is being measured and pH was enabled), alarm 2
setpoint (no units), alarm 3 setpoint (no units),
output 2, or blank
00 (lightest)-99 (darkest); the display contrast
changes as the number changes
Timeout returns the display from any other screen
to the main display if no key is pressed before the
timeout value is exceeded.
Identifies controller in multi-drop HART applications.
SECURITY CAUTION
SECTION 5.0
and
keys
29