Download Print this page

KERN KFB-TAM Operating And Installation Instructions page 117

Display device

Advertisement

Available languages

Available languages

10 Data output RS 232C
You can print weighing data automatically via the RS 232C interface or manually by
pressing
via the interface according to the setting in the menu.
This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communication between
the weighing system and the printer.
• Use a suitable cable to connect the display unit to the interface of the printer.
Faultless operation requires an adequate KERN interface cable.
• Communication parameters (baud rate, bits and parity) of display unit and
printer must match. For a detailed description of interface parameters see
chap. 8, menu block „P1 COM" or ‚"P2 COM"
10.1 Technical data
Connection
9 pin d-subminiature bushing
Baud rate
Optional 600/1200/2400/4800/9600
Parity
8 bits, no parity / 7 bits, even parity / 7 bits, odd parity
KFB/KFN-TAM-BA_IA-e-1832
KFB-TAM
Pin 2 RXD
Pin 3 TXD
Pin 5 GND
KFN-TAM
Pin 2 TXD
Pin 3 RXD
Pin 5 GND
49

Advertisement

loading

This manual is also suitable for:

Kfn-tam