8. Adjustment

Switch-off balance and attach a carrying help if necessary.
Switch-on balance with attached carrying help and during the selftest
press the TARE button. "P1 - - -" will be displayed.
Enter password "000" :
Select a number by the M+ key
Increase the numeric value using the TARE key
Confirm by the ZERO key. The first function „F0 cal" is displayed.
Press the ZERO button, „UnLoAD" will be displayed.
For verified models press the adjustment button at the balance bottom
Unload the balance and wait until the [] above
Press the ZERO button, the currently set adjustment weight is displayed.
In order to change, select the number to be changed by the M+ button and
set the desired value using the TARE button, the active digit flashes.
Confirm by the ZERO button, „LoAd" will be displayed.
Attach the adjustment weight and wait until the [] above
Press the ZERO button.
After successful adjustment the balance carries out a selftest, then it automatically
returns to weighing mode.
An adjusting error or incorrect adjusting weight will be indicated by the error
message; repeat adjustment procedure