10. Error messages

Error
Description
message
Err 1
Wrong date
Err 2
Wrong time
Err 4
Error during zero setting
Err 5
Keyboard error
Value outside the A/D changer
Err 6
range
Error "Determination of
Err 7
percentage"
Err 8
Incorrect adjusting weight
The displayed weight is
Err 9
permanently changing
Err 10
No WLAN connection
Error "Communication log"
Err 11
Error "Totalizing"
Err 12
Error "Gravitational constant"
Err 15
"Taring" error
Err 17
Unable to initialise zero point
Err 19
- - ol - -
Maximum load exceeded
--lo--
Underload
Adjustment error
Fai l h / fai l l
/ fai l
Capacity of rechargeable battery
Ba lo / lo ba
exhausted
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
34
Possible causes / Elimination
Enter date in format "yy;mm;dd", see
7
chap.
"F7 da"
Enter time in format "hh;mm;ss", see
7
chap.
"F7 da"
Zero range exceeded
Check if the balance is not loaded
Improper operation of the balance
Damaged weighing cell
Damaged electronics
Increase the value to >0.5 d
Check value for adjustment weight, see
chap. 1
Draught/air movement
Table/floor vibrations
Weighing pan has contact with other
objects.
Check menu setting
"F5 commodewifi"
Check the communication settings
Number of totalizing operations > 99
Capacity of balance exhausted
Value out of range 09.xx -1.0xx
Taring range exceed or not achieved
Measuring cell defective / overloaded
Object on weighing pan / contact
Main board defective
Adjustment required
Reduce load
Check whether the balance has been
damaged
Negative weight, check platform and
restart or adjust.
Check value for adjustment weight, see
chap. 1
Repeat adjustment process
Recharge battery
HFD-BA-e-1820