Inserting The Specimen Stage, Orientable, With Ball-Joint - Leica CM3600 XP Instruction Manual

Cryomacrotome
Hide thumbs Also See for CM3600 XP:
Table of Contents

Advertisement

7.
Operating the Leica CM3600 XP

7.5.1 Inserting the specimen stage, orientable, with ball-joint

92
To insert the specimen stage, orientable, with ball-and-socket joint
proceed as described below. The special holder, orientable, with ball
joint is inserted the same way.
Tools required:
• Fork wrench (size 17).
Proceed as follows:
1. Use the knee lever to move the sledge to the right-side reversal point.
For further information, refer to chapters 3.2.2 and 6.4.9.
2. Comply with the safety instructions provided in chapter 7.5.
3. Open the window. For further information, refer to chapter 3.4.3.
4. Place specimen stage (1) on T-piece (2).
5 Rotate specimen stage (1) 90° clockwise.
6. Orient the specimen stage (1) by eye.
7. Tighten screw (3).
8. Check specimen stage (1) for correct orientation and proper tighten-
ing.
Instruction manual V 6.1 - 07/2007