OPERATION
5
Once a flash drive with valid files are found, select the desired range to update using the
buttons.
While the system is updating, the word "busy" appears as it performs a self-check of the
software installation.
If the device finds a flash drive with a previous version of software it will show the word FILE. You
will need an updated file on the flash drive.
Once installed the system will automatically turn off and restart. The entire process takes about
60 seconds to complete.
INFORMATION!
Be certain to check the configuration as changes may have taken place either with new software
or a new range.

5.10 Reset to factory calibration

CAUTION!
Make sure that only authorised and trained personnel perform changes to the calibration.
INFORMATION!
Incorrect settings in the parameters may result in the output of incorrect measured values and
switching points.
Proceed as follows to reset the device to the factory calibration and delete possible user
calibrations:
1 Press the ^ and the
50
button for 2 seconds to get access to the reset function.
www.krohne.com
OPTISYS TUR 1060
01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en