KROHNE OPTISYS TUR 1060 Handbook

KROHNE OPTISYS TUR 1060 Handbook

Compact measuring system for turbidity
Table of Contents

Advertisement

Quick Links

OPTISYS TUR 1060
OPTISYS TUR 1060
OPTISYS TUR 1060
OPTISYS TUR 1060
Compact measuring system for turbidity
© KROHNE 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en
Handbook
Handbook
Handbook
Handbook

Advertisement

Table of Contents
loading

Summary of Contents for KROHNE OPTISYS TUR 1060

  • Page 1 OPTISYS TUR 1060 OPTISYS TUR 1060 OPTISYS TUR 1060 OPTISYS TUR 1060 Handbook Handbook Handbook Handbook Compact measuring system for turbidity © KROHNE 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 2 All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH. Subject to change without notice. Copyright 2019 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany) www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 3: Table Of Contents

    4.5.1 Connecting the cables in the junction box power assembly ..........23 4.5.2 Alarms terminal (signal output) ................... 24 4.5.3 RS 485 / 4...20 mA signal output................... 25 4.5.4 Connecting the sensor interconnect cable................26 5 Operation 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 4 6.6 Returning the device to the manufacturer..............56 6.6.1 General information......................56 6.6.2 Form (for copying) to accompany a returned device............57 6.7 Disposal .......................... 57 7 Technical data 7.1 Measuring principle......................58 www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 5 CONTENTS OPTISYS TUR 1060 7.2 Technical data table ....................... 59 7.3 Dimensions and weight ....................61 8 Notes 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 6: Safety Instructions

    OPTISYS TUR 1060 1.1 Intended use The OPTISYS TUR 1060 is available in white light and infrared versions. Both versions are designed to measure online the turbidity of water. Furthermore the two following different measuring ranges are available: 0...100 NTU/FNU or 0...1000 NTU/FNU 1.2 Certification...
  • Page 7: Safety Instructions From The Manufacturer

    The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 8: Product Liability And Warranty

    This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of icons as shown below. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 9: Warnings And Symbols Used

    In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 10: Device Description

    2 Tubing kit (drain vent, shut-off clamp, seal screw, backpressure valve, connection tubes with fittings for ultrasonic cuvette assembly) 3 Junction box power assembly 4 Desiccant pack (desiccant pouch with humidity indicator) 5 Documentation www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 11: Device Description

    10 Drain hose 11 Terminal box 12 Intake hose 13 Intake hose connection (1/4 inch OD flexible tubing), has to supply the sensor with a dependable sample flow 14 Pressure regulator 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 12 Liquid crystal display (LCD) For a better readability in low light or no light conditions the LCD is backlight. The backlight always stays on, but the brightness is adjustable. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 13: Nameplates

    2 Observe the operation and installation instruction, Electronic / electric device waste marking, China RoHs 3 Electrical data 4 Protection category 5 Serial number 6 Order code 7 Device type 8 Part number 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 14: Installation

    • Input pressure range: 0.01...7 bar / 14.5...101.5 psi (built in regulator set at 1 bar / 15 psi) • Maximum allowable flow rate of the cuvette: 0.1...1,5 l/min / 0.026...0.4 gal/min • Maximum fluid temperature: +50°C / +122°F www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 15: Installation Order

    • Prior to installing the desiccant pouch for the first time, remove the shipping support (plastic tube with a red flag reaching outside) within the upper part of the device; after removing you can discard the tube. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 16: Selecting Site And Mounting

    • Refer to chapter "Dimensions" for the device dimensions. • Use screws M6 / 1/4" to fix the electronic device 2 and M4 / 3/16" to fix the junction box power assembly 3. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 17: Mounting Dimensions

    Leave at least 20 cm / 8" of free space above the device for easy service (e.g. removal of the flow head and insertion of calibration cuvettes). [mm] ["] 5.79 3.54 3.43 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 18: Connecting The Hoses

    • Opaque hose material to prevent algae growth if direct sunlight can reach the hose. INFORMATION! To prevent that the drain water causes any damages, assure that the drain hose leads to a suitable drain site. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 19: Water Intake

    Slide the hose into the Quick Connect fitting on the right-hand side and check that the hose is completely pressed against the stop. Pull carefully but firmly on the pipe to check if it is secured. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 20: Inserting And Fixing The Ultrasonic Cuvette With Flow-Through Unit

    The device can only detect a new cuvette if it works in the normal operation mode ("AUTO"). If the device operates correctly after inserting a new cuvette, "AUTO" will flash on the display. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 21: Electrical Connections

    All of the electrical connections to the device run via the junction box power assembly. For shipment the manufacturer inserts plugs into the alarm and the 4...20 mA/RS 485 cable gland to waterproof the device. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 22: Circuit Breaker And Specifications Of Power Supply

    Before fixing a cable, look on the labels on the board and especially regard polarities! • To establish the electrical connections in the correct way, follow the order of the following sections and their instructions. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 23: Connecting The Cables In The Junction Box Power Assembly

    • Use the strain relief strap to reduce the tension on the power terminals 3. • Close the terminal box and assure that it is properly sealed. Figure 4-2: Connecting the cables in the junction box 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 24: Alarms Terminal (Signal Output)

    For detailed information concerning the configuration of the alarms refer to Configuring the alarm relays on page 41. Connections are labeled beneath the large terminal block. These connections are on the upper row. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 25: Rs 485 / 4

    • Fix the cables on the terminal as shown in the drawing 2. to 4 • Use the strain relief strap to reduce the tension on the power terminals. • Close the terminal box and assure that it is properly sealed. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 26: Connecting The Sensor Interconnect Cable

    Figure 4-4: Connecting cable www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 27: Operation

    4...20 mA output or the digital RS 485 output, depending on the options selected. • Initial seetings like Date and Time must be set now. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 28: Display And Operating Keys

    During normal operation, the display shows an arrow beside "AUTO" in the upper left corner. Simultaneously the lower row displays the current scale and the upper row the measured reading. The following drawing is an example for a display during normal operation: www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 29: Getting Access With Activated Security Access Function

    If you have entered the valid code, the main menu with the arrow beside "CAL" appears. If you have entered a wrong code, it returns to the normal measuring mode ("AUTO"). 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 30: Change The Security Access Code

    Now you can change the code as desired. • Use to select the number and use to select the digit. • If the code is adjusted, press ^ to get back in the normal measuring mode. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 31: Menu Topology

    2 x = CONFIG, 3 x HOLD). Press ^ to enter one of the 4 main menus. • Press to move through the submenu items. • Press to change the parameter in the submenu. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 32: Menu Mode Structure

    CAL - Calibrate Calibration starts (wait ^ ^ ^ ^ 30 seconds) Calibration starts (wait 60 seconds) 0.02 Calibration starts (wait 30 seconds) ABRT Abort the calibration www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 33 OFST OFST - Offset OFST OFST CODE CODE - Access CODE CODE Code RESP RESP - Signal RESP RESP 1...60 Averaging RES - Display 0.0000 Resolution 00.000 000.00 0000.0 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 34 4MA - 4 mA Adjust -200...200 20MA 20MA - 20 mA 20MA 20MA -1000...1000 Adjust ID ID ID ID - Instrument ID 0001...9999 INVL - Logging INVL 1...60 INVL INVL Interval www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 35: Calibration Mode (Main Menu)

    The EPA and ISO recommend that on-line turbidimeters be calibrated with a primary standard at least once every three months if they are to be used for reporting purposes. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 36: Calibration Standards

    The arrow in the display appears beside "CAL", the lower display line shows alternating "100" (turbidity value of the first calibration liquid in NTU/FNU) and ^, the upper line shows the real-time reading. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 37 • Repeat the procedure above and finally press ^ to accept the 0.02 NTU/FNU calibration. The lower line of the display counts down the progress of the calibration step, afterwards the device returns to the normal operation mode ("AUTO"). 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 38: Calibration Procedure (Offset Calibration)

    The offset calibration is complete and the device goes to the next submenu in the configuration main menu. • If you want to return to the normal measuring mode, press ^. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 39: Calibration Error

    The device has the ability to customise it according to your specific needs any time during normal operation with the help of the configuration mode. Be aware of the following device properties: 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 40: Setting The 4

    • Use the buttons to define the upper turbidity limit which is assigned to the 20 mA output level and press The device switches to the next menu item, i.e. "Error level". www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 41: Baud Rate

    The device has two alarm relays that operate as two independent programmable alarms. To fully program each alarm, you have to input the work mode of the alarm function, the alarm set point and the delay time for the alarm. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 42 If you choose "HI", "LO" or "Error" a prompt appears to set up the alarm set point. AUTO CONFIG • Use the buttons to select the desired alarm set point in the upper row. • Accept your selection by pressing www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 43: Offset Calibration

    Execute the following steps to change the setting of the function: • Use the buttons to enable (upper row: ON) or disable (upper row: OFF) the security access or leave the setting 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 44: Speed Of Response (Signal Averaging)

    Changing a setting and switching to the next item • Use the buttons to change the setting. • Accept your selection by pressing The device switches to the next item of the menu. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 45: Display Backlight Brightness

    The device switches to the next item of the menu. The most common unit of the measured value and the default setting is NTU (Nephelometric Turbidity Units), but FNU (Formazin Nephelometric Units) is also available. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 46: Ultrasonic Cleaning

    The device switches to the next item of the menu. • Use to select the desired number of bits or leave the setting, accept your selection by pressing The device switches to the next item of the menu. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 47: Ma And 20 Ma Adjustment

    The flashing digit is the number that is currently being adjusted. • Accept your selection by pressing The device switches to the next item of the menu. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 48: Logging Interval

    The MTOL+ has a displayed date (month / day / year) and time (hour / minute / Daylight Savings) that must be set upon initial installation/setup. The next five menus are used for this setup. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 49: Daylight Saving Time

    If either a flash drive is not installed or a newer version of software is not detected this error screen will appear. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 50: Reset To Factory Calibration

    Proceed as follows to reset the device to the factory calibration and delete possible user calibrations: 1 Press the ^ and the button for 2 seconds to get access to the reset function. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 51: Faults: Reasons And Remedies

    These failures consist of failures in the CPU, A/D, EEPROM or other devices internal to the instrument. The word FAIL will display on the lower row. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 52 Check for leakages and replace devices if cuvette. necessary. Improper calibration. Recalibrate the device (details on page 35). Measuring result is Improper calibration. Recalibrate the device (details on page 35). lower than expected www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 53: Service

    • Reassemble everything in reverse order. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 54: Replacing Desiccant Pouch And Humidity Indicator

    To prevent the device from damages, do not try to replace the lamp on your own! If you need to replace a lamp, contact the manufacturer or your local sales office. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 55: Availability Of Services

    Grab sample cuvette, 10 pack XGAM 010080 Solid validation standard (cuvette) XGAM 010100 Calibration kit (0.02, 10, 100 NTU/FNU) XGAM 010150 Calibration kit (0.02, 10, 1000 NTU/FNU) XGAM 010160 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 56: Returning The Device To The Manufacturer

    • such dangerous substances, to enclose a certificate with the device confirming that it is safe to handle and stating the • product used. www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 57: Form (For Copying) To Accompany A Returned Device

    The user must dispose of the WEEE to a designated collection point for the recycling of WEEE or send them back to our local organisation or authorised representative. 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 58: Technical Data

    • FNU (Formazin Nephelometric Unit) Figure 7-1: Measuring principle for turbidity measurement 1 Light source 2 Reference sensor 3 Sensor 4 Reflected light beam 5 Particles 6 Emitted light beam www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 59 0...1000 NTU/FNU range) > 40 NTU/FNU: ±5% of reading (0...100 or 0...1000 NTU/FNU range) ±2% of reading or ±0.02 NTU/FNU (0...10 NTU/FNU range) Resolution 0.0001 NTU/FNU (below 10 NTU/FNU) selectable 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 60 This device fulfils the statutory requirements of the EC directives. The manufacturer certifies successful testing of the product by applying the CE mark. Listed to UL 61010-1 (Ed.3) and certified to CAN/CSA 22.2 No. 61010-1-12 (R2017) www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 61 TECHNICAL DATA OPTISYS TUR 1060 7.3 Dimensions and weight Dimensions [mm] ["] 8.19 88.77 1.18 3.43 5.83 133.71 5.26 13.79 2.44 Shipping weight: 2.8 kg / 6.17lbs 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 62 NOTES OPTISYS TUR 1060 www.krohne.com 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en...
  • Page 63 NOTES OPTISYS TUR 1060 01/2019 - 4007152501 - MA OPTISYS TUR 1060 R01 en www.krohne.com...
  • Page 64 • Process Analysis • Services Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 47058 Duisburg (Germany) Tel.: +49 203 301 0 Fax: +49 203 301 10389 info@krohne.com The current list of all KROHNE contacts and addresses can be found at: www.krohne.com...

Table of Contents