APPENDIX B
Appendix B
Variable Definitions
From the VARIABLES page, press the
variable required. Press the
variable using the
the new option. Scroll down to the next variable required. To return to the MAIN Page
MENU/MAIN
press the
UNITS (single channel), CHN1, CHN2
Variable:
Adjustable Range/Options:
Description: The moisture units for Channels 1 and 2.
apply
Variable:
Adjustable Range/Options:
Description: The pressure value entered here is the pressure value at which dew point
Variable:
alarm
Adjustable Range/Options:
Description: The set point for the process alarm that activates the channel alarm relay
OP1MIN (single channel), OP3MIN
Variable:
Adjustable Range/Options:
Description: The minimum moisture value that 4 mA represents for the respective
42
SELECT
Up
() or
Down (
button.
Ppm(v) IG, DP@PR idl Gas, DP@PR Nat Gas, mgm-3 Nat
Gas, Ppm(v) NG, LBMMSCF, Dew point
i.
Dew point can be displayed in either °C or °F whether it is at
the sample pressure or if it is calculated to another.
ii.
When calculating the dew point at another pressure (
PR
apply
) the
option below is enabled to allow the entry of
the pressure at which the dew point is to be calculated at.
iii.
mA1 and mA3 outputs represent the setting of Channel 1
and 2 respectively.
NOTE: Changing the units clears the data logging memory and resets the
analog outputs and alarm settings.
Pressure value in the units of pressure
will be calculated at, if
the unit of moisture measurement. If any other unit is selected then the
option will show
n/a
.
-100 to +100°C, 0 to 1200 LBMMSCF, 0.0 to 3000.0
Ppm(v), 0 to 1000mgm-3
when the moisture value exceeds it.
-100 to +100°C, 0 to 1200 LBMMSCF, 0.0 to 3000.0
Ppm(v), 0 to 1000mgm-3
output.
Promet EExd User's Manual
Up
() or
Down (
button to enter the range/options area. Change the
) buttons and press the
DP@PR Nat Gas
or
DP@PR IDL Gas
97090 Issue 16.1, November 2019
) buttons to highlight the
SELECT
button to set
DP @
is selected as
Need help?
Do you have a question about the Promet EExd and is the answer not in the manual?