MEDRAD
Stellant CT Injection System with Certegra
Frequency: Daily and In the Case of Visible Contamination
Materials:
NOTE: If in a country where Sani-Cloth

15.2.2 Disinfecting the Injector Head

Frequency: Daily and In the Case of Visible Contamination
Material(s):
15 - 68
Clean, soft, non-linting wipe
Warm water
Cleaning Agent:
Use either Sani-Cloth
CaviWipes™ Disinfecting Towelettes (EPA registration # 46781-8) within the United
States except where noted.
Use either Sani-Cloth
Towelettes in all other countries except where noted. If neither Sani-Cloth
cidal Disposable Cloth nor CaviWipes™ Disinfecting Towelettes are available within
your country, then use a disinfectant with 0.25% - 0.30% quaternary ammonium com-
pounds and 14.8% - 20% isopropanol.
1. Turn off power to the injector.
2. Clean the injector head surfaces (excluding pistons and syringe interface, refer to Figure 15 -
3: Piston) with a clean, soft, non-linting wipe dampened with warm water (wet, but not drip-
ping) for a minimum of 1 minute and until visibly clean. Replace wipe if wipe becomes visibly
soiled. During the 1 minute cleaning time, make certain that all seams, recessed areas, and
buttons are clean. Refer to
pistons and syringe interface.
3. Thoroughly dry the injector head using a dry, clean, soft, non-linting wipe.
4. Wipe all external surfaces with the exception of the syringe pistons and syringe interface
(refer to Figure 15 - 3: Piston) with Sani-Cloth
ibly clean. Make certain that all seams, recessed areas, and buttons are clean. Refer to
"Cleaning the Pistons and Syringe Interface"
face.
equivalent cleaning agent (refer to the materials listed above), follow the same instructions
specified in step 4.
5. Allow the injector head to thoroughly air dry.
6. Inspect the injector head to ensure all surfaces are clean.
7. If any visible contrast media or other debris is evident, repeat cleaning instructions steps 2
through 6 until no visible soil is detected.
Disinfecting Agent:
Use either Sani-Cloth
CaviWipes™ Disinfecting Towelettes (EPA registration # 46781-8) within the United
States except where noted.
Use either Sani-Cloth
Towelettes in all other countries except where noted. If neither Sani-Cloth
cidal Disposable Cloth nor CaviWipes™ Disinfecting Towelettes are available within
your country, then use a disinfectant with 0.25% - 0.30% quaternary ammonium com-
pounds and 14.8% - 20% isopropanol.
®
Workstation
®
Plus Germicidal Disposable Cloth (EPA registration # 9480-6) or
®
Plus Germicidal Disposable Cloth or CaviWipes™ Disinfecting
"Cleaning the Pistons and Syringe Interface"
®
Plus or CaviWipes™ for 1 minute or until vis-
to clean the syringe pistons and syringe inter-
®
Plus or CaviWipes™ are not available, and using an
®
Plus Germicidal Disposable Cloth (EPA registration # 9480-6) or
®
Plus Germicidal Disposable Cloth or CaviWipes™ Disinfecting
®
Plus Germi-
to clean the syringe
®