Thermo Scientific Sorvall X Pro Series Instructions For Use Manual page 93

Table of Contents

Advertisement

CAUTI ON
Drive and door lock can be damaged by entering liquids. Do not allow
liquids, especially organic solvents, to get on the drive shaft, the drive
bearings or the centrifuge door locks. Organic solvents break down the
grease in the motor bearing. The drive shaft could lock up.
Touchscreen
1.
Pull out the power supply plug.
2.
Clean the touchscreen using a dry microfiber cloth.
3.
If necessary moisten the microfiber cloth and wipe the touchscreen again.
Ventilation Grid
To clean the ventilation grid proceed as follows:
1.
Pull out the power supply plug.
1.
Remove the 2 screws of the ventilation grid ➀ on the right side of the centrifuge.
2.
Remove the ventilation grid ➁ by pushing it downards.
3.
Use a vacuum cleaner to clean the ventilation grid and if necessary the condenser. Use a soft brush for
detailed cleaning if needed.
4.
Reinstall the ventilation grid.
Figure 5–1: Removing the ventilation grid
CAUTI ON
Cutting injuries cann occur due to sharp metal. Do not touch the condenser
with your hands when the ventilation grid is removed.
Screws
Ventilation Grid
5-3
Maintenance and Care