Thermo Scientific Sorvall X Pro Series Instructions For Use Manual page 71

Table of Contents

Advertisement

NOTICE The AM/PM division does not appear when the radio button in the "Settings -> Display -> Time"
screen is set to 24hr (see
7.
Tap the Time On field and enter the time of day at which you wish the centrifuge to start.
8.
Tap the Temp field (refrigerated models only) and set the target temperature for pre-tempering, if desired.
NOTICE Observe the note about cooling and take precautions to ensure that the lid is closed at the
specified start time.
9.
Tap more days and repeat this procedure for as many days of the week as required.
10. Tap the Save button to save the new "Auto On" schedule for the centrifuge.
A pop-up window appears, alerting you to the risk of condensation pools freezing over in the centrifugation
chamber.
NOTICE Observe the condensation warning and take precautions to prevent condensation from freezing
in the centrifugation chamber.
11. To confirm the changes, tap the OK button in the pop-up window shown above.
You will be returned to the "Schedule" screen.
12. On the "Schedule" screen, tap the slider above the schedule field, so it changes to Enabled.
Your "Auto On" schedule is active now, the centrifuge will turn on automatically at the specified times.
Auto Off Scheduling
1.
Tap the Settings icon
2.
Tap the Controls button
The "Controls" screen appears.
Tap the Scheduling button on the "Controls" screen.
3.
The "Scheduling" screen appears.
If you have not configured any schedules yet, both the "Auto On" and "Auto Off" schedules will be disabled.
4.
Tap the Edit button for the "Auto Off" option to start setting up a schedule.
The "Edit Auto Off" screen appears.
5.
Choose the day(s) of the week and set the stop times as described further above for "Auto On".
6.
Tap the Auto open lid checkbox to have the lid open automatically after the centrifugation run, if desired.
NOTICE Observe the notes about how an open lid interferes with "Auto On" and about the requirement
to open the lid along with "Auto Off" after cooling to prevent condensation.
7.
Tap more days and repeat this procedure for as many days of the week as required.
8.
Tap the Save button to save the new "Auto On" schedule for the centrifuge.
If you have chosen the "Auto Open Lid" option, a pop-up window appears to alert you to the risk that an
open lid interferes with the next "Auto On" schedule.
NOTICE Observe the warning about the open lid and take precautions to have the lid closed before the
next "Auto Off" schedule is effective.
9.
To confirm the changes, tap the OK button in the pop-up window shown above.
You will be returned to the "Schedule" screen.
10. On the "Schedule" screen, tap the slider above the schedule field, so it changes to Enabled.
Your auto-off schedule is active now. The centrifuge will turn off automatically at the specified times.
The two sliders above the "Auto On" and "Auto Off" fields read "Enabled" (see
"Time" on page 3-43
for details).
on the navigation bar.
on the "Settings" screen.
3-39
Graphical User Interface
Figure
3–63).