Battery Indicator And Battery Change; Low Battery; Flat Battery - Replacement; Which Batteries And Which Procedure - Microlife BP B1 Classic Manual

Hide thumbs Also See for BP B1 Classic:
Table of Contents

Advertisement

Available languages

Available languages

switched off beforehand) until «CL ALL» appears and then
release the button. To permanently clear the memory, press the
time button while «CL ALL» is flashing. Individual values cannot
be cleared.
Cancel deletion: press ON/OFF button 1 while «CL
ALL» is flashing.

5. Battery Indicator and Battery change

Low battery

When the batteries are approximately ¾ empty the battery symbol
AM will flash as soon as the device is switched on (partly filled
battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.
Flat battery – replacement
When the batteries are flat, the battery symbol AM will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment 6 at the back of the device.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in
«Section 1.».
The measurements stored in the memory are deleted when
the batteries are removed from the battery compartment
(e.g. when replacing batteries).

Which batteries and which procedure?

Use 4 new, long-life 1.5 V, size AAA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.
6. Error Messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
Error
Description Potential cause and remedy
«Err 1»
Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat the
AR
measurement.*
4
Error
Description Potential cause and remedy
«Err 2»
Error signal During the measurement, error signals
were detected by the cuff, caused for
AQ-B
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
Abnormal
An adequate pressure cannot be gener-
cuff pres-
ated in the cuff. A leak may have
AQ-C
sure
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«Err 5»
Abnormal
The measuring signals are inaccurate
result
and no result can therefore be displayed.
Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«HI»
Pulse or
The pressure in the cuff is too high (over
cuff pres-
299 mmHg) OR the pulse is too high
sure too
(over 200 beats per minute). Relax for 5
high
minutes and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measurement.*
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
7. Safety, Care, Accuracy Test and Disposal
Safety and protection
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section.
 Protect it from:
- water and moisture
- extreme temperatures