Maintenance; Draining The Bath (Where Applicable) - Thermo Scientific MaxQ 7000 User Manual

Benchtop water bath shakers
Table of Contents

Advertisement

Draining the Bath
(where applicable)
Thermo Scientific

Maintenance

Section 7
Note Make no attempt to service or repair a Thermo Scientific product
under warranty before consulting your Thermo Scientific dealer. After the
warranty period, such consultation is still advised, especially when the
repair may be technically sophisticated or difficult.
Warning Disconnect plug from electrical outlet before attempting any
maintenance or repair of the unit. Any internal adjustments or repairs
must be performed by a qualified service representative.
Wash the exterior of the unit with a soft cloth using a solution of mild
soap and water, rinse off with distilled water and dry thoroughly.
Wash bath and components, suspension arms, platform and splash baffles
with a solution of mild soapy water. Use a soft cloth, sponge or synthetic
sponge scrubber. Rinse all components thoroughly and dry completely.
DO NOT USE chlorine bleach, chlorine-based cleanser, abrasives,
ammonia, steel wool or scouring pads with metallic content. These will
damage #302 stainless steel. If metal parts drop into the bath, remove
them at once. They can leave rust deposits on the stainless steel surface
that may be difficult to remove.
1. Disconnect shaker from electrical power.
2. Attach one end of a suitable length of 1/2" (0.64cm) tubing to the
barbed end of the adapter fitting supplied with the unit.
3. Attach the opposite end of the adapter fitting to the quick-connect on
the front of the shaker.
4. Place the other end of the tubing in a sink, floor drain or in an
adequately sized receptacle. Bath will drain when quick-connect reset
button is pressed.
s
MaxQ 7000
7-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Shka7000Shka7000-1ceShke7000Shke7000-1ce

Table of Contents