Model number
Speed
Speed Shut off
Timer
Unbalanced Load
Thermo Scientific
SHKA7000
None
None
Continuous operation (hold)
or timed operation from 1 to
60 minutes
None
Optional Platform Dimensions - in. (cm)
Catalog Number L x W
30105
SHKA7000-1CE
None
None
Continuous operation (hold)
or timed operation from 1 to
60 minutes
None
Universal 13" x 11" (33 x 27.9 cm)
SHKE7000
Audible with flashing LED
indicate when speed devi-
ates more than 10% of set
point.
When speed deviates 10%
of set point, unit will shut
down immediately.
Beeps twice when time has
expired. Shaking motion
stops.
If the unit is running in an
unbalanced condition, an
alarm will sound and the
shaker will stop until the
end user corrects the condi-
tion. The speed display will
flash "bAL" on speed panel
LED.
Section 2
Specifications
SHKE7000-1CE
Audible with flashing LED
indicate when speed devi-
ates more than 10% of set
point.
When speed deviates 10%
of set point, unit will shut
down immediately.
Beeps twice when time has
expired. Shaking motion
stops.
If the unit is running in an
unbalanced condition, an
alarm will sound and the
shaker will stop until the
end user corrects the condi-
tion. The speed display will
flash "bAL" on speed panel
LED.