7
Function group "extended calibr." (05)
7.1
Function "selection" (050)
Select the function of the extended calibration.
Selection:
• common (e.g. "Level correction", "Output damping", etc.)
• mapping
• delete mapping
7.2
Function "check distance" (051)
This function triggers the mapping of interference echoes. To do so, the measured distance must be
compared with the actual distance to the product surface. The following options are available for
selection:
Selection:
• distance = ok
• dist. too small
• dist. too big
• dist. unknown
• manual
• probe free
distance too small
+
7 Function group "extended calibr." (05)
distance = ok
+