Levelflex M
!
Endress+Hauser
FMP43:
ε
Media group
DC (
r)
1
1.4 to 1.6
2
1.6 to 1.9
3
1.9 to 2.5
4
2.5 to 4
5
4 to 7
6
> 7
FMP41C, FMP45:
ε
Media group
DC (
r)
1
1.4...1.6
2
1.6...1.9
3
1.9...2.5
4
2.5...4
5
4...7
6
> 7
The lower group applies to very loose or loosened bulk solids.
Reduction of the max. possible measuring range by means of:
• extremely loose surfaces of bulk solids, e.g. bulk solids with low piled density when filled pneu-
matically.
• Build-up, primarily of moist products.
Note!
Due to the high diffusion rate of ammonia it is recommended to use the FMP45 with gas-tight bush-
ing for measurements in this medium.
Typical liquids
Condensed gases, e.g. N
, CO
2
2
Liquefied gas, e.g. propane
Solvent
Frigen / Freon
Palm oil
Mineral oils, fuels
Benzene, styrene, toluene
Furan
Naphthalene
Chlorobenzene, chloroform
Cellulose spray
Isocyanate, aniline
Aqueous solutions
Alcohols
Acids, alkalis
Typical liquids
Condensed gases, e.g. N
, CO
2
Liquefied gas, e.g. Propane
Solvent
Frigen / Freon
Palm oil
Mineral oils, fuels
Benzene, styrene, toluene
Furan
Naphthalene
Chlorobenzene, chloroform
Cellulose spray
Isocyanate, aniline
Aqueous solutions
Alcohols
Acids, alkalis
3 Function group "basic setup" (00)
Typ. measuring range
—
4 m (354")
Typ. measuring range
4 m (157"), when installed in metallic pipes
2
9 m (354")
12 m (472")
16 m (629")
25 m (984")
30 m (1181")
17