Advertisement

4.1.3.3. Timing Setting
1.
X ray generator is ready for X ray shooting and begins to release X ray.
2.
Workstation receives "Exposure Prohibited" from Panel.
3.
Panel starts uploading Pre-dark image and Light image to Workstation for preview. If hardware
offset is selected, panel would do offset first, and then upload preview image (2X2 binning).
4.
Panel starts uploading Post-dark image to Workstation. If hardware offset is chosen, panel would
do correction and calibration first, then upload processed image to Workstation.
5.
Workstation receives "Exposure Enable" from Panel.

4.2. Connection Build

Open
SDK
and
product start connection
Confirm the IP address and
the Port are the same as the
value in config.ini.
The port should use the
default value of 28000
INC.................................................................................................................................
choose
Operation
KONICA MINOLTA,
...............................53