SKR 4000 DIRECT DIGITIZER User Manual
If
the
value
meet
expected
value,
"Accept", then get the other 7
images.
If the value does not meet the
expected value, please do not
click
the
"Accept",
adjust
the
exposure
dose ,then click the "PREP"
to get light image again
After getting 8 images, click
"Generate" to generate gain
template
Note: 1 please use software post offset mode.
2. Make sure your x ray dose is right, if your dose is out of the range, IDetector will remind you to
adjust the dose. Then you can click "start creating" and try again.
3. If users operate with two panels, SDK has a probability of quit automatically.
4.5. Image Check and upload
"OPEN" provides two features for image check and uploading. Local Image Check, Panel Image
Upload. Local Image Check defines function to check image saved in Workstation. Panel Image Upload
defines function to upload images stored in panel.
4.5.1. Local Image Check
Click "Local File" button in
"Local File" UI, choose the
specified file
Choose
images
stored
60
..................................................................................................................................................KONICA
MINOLTA, INC.
the
click
and
in /