Calibration Information

Maintenance

Maintenance of the Meter consists of battery and fuse replacement, as well as case cleaning.

Fuse Test (115, 117)

To test the fuse:
1. Set the rotary switch to Ω.
2. Plug a test lead into the v jack and touch the probe to the 10A jack, as shown in Figure 1.
If the display shows a resistance value in the range of that shown in Figure 1, the fuse is good.
If the display reads 0L, replace the fuse and test again.
If the display shows any other value, have the Meter serviced. See How to Contact Fluke earlier in this
document.
2
Figure 1. Test the Fuse
<.5
OK
X
OK
OK
OK