Fluke 110 Calibration Information Manual page 22

True-rms multimeter
Table of Contents

Advertisement

110/113/114/115/116/117
Calibration Information
Table 5. Calibration Adjustment Steps (110/114/115/116/117) (cont.)
Rotary
Switch
110/114
Position
Set calibrator to standby, reconfigure leads, and program for amps output.
NA
NA
NA
$
μ
DC
Amps
NA
[1]
Models listed in this column also refer to the "C" version of the model. For example, model 115 steps are valid for the 115C.
[2]
Do not calibrate the 117 or 117C with a line-frequency power source nearby (for example, fluorescent light or power strip). These
devices can produce errors in the VoltAlert calibration.
[3]
Wait an additional 5 seconds after calibrator has settled before pressing g.
Rotary Switch
Position
k
CHEK
mV ac/dc
Ω
Ohms
k
CHEK
V ac
[1]
To keep from tripping the calibrator to standby, ramp up the voltage in 50 V increments with a 5 second delay between increments.
18
Display Reading
[1]
[1]
115
116
NA
C022
NA
C023
NA
C022
NA
C023
Table 6. Calibration Adjustment Steps (113)
Calibration Steps
C/01
C/02
C/03
C/04
C/05
C/06
C/07
C/08
C/09
C/10
C/11
C/12
C/13
C/14
Input
Terminals
[1, 2]
117
A and COM
C022
A and COM
C023
NA
+ and COM
NA
+ and COM
Input Terminals
+ and COM
0 V, 0 Hz
+ and COM
300 mV, 0 Hz
+ and COM
100 mV, 0 Hz
+ and COM
-300 mV, 0 Hz
+ and COM
60 mV, 0 Hz
+ and COM
600 mV, 0 Hz
+ and COM
600 mV, 60 Hz
+ and COM
600 e, 2-wire comp
+ and COM
6 ke
+ and COM
60 ke
+ and COM
600 ke
+ and COM
6 V, 60 Hz
+ and COM
60 V, 60 Hz
+ and COM
600 V, 60 Hz
Calibrator
Source Value
6 A, 60 Hz
[3]
6 A, 0 Hz
μ
600
A, 60 Hz
μ
600
A, 0 Hz
Calibrator Source Value
[1]

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

113114115116117

Table of Contents