Fluke 110 Calibration Information Manual page 13

True-rms multimeter
Table of Contents

Advertisement

Table 3. DMM Performance Tests: 110/114/115/116/117 (cont.)
Step
Function
Range
9.
600
R
Continuity
10.
600
11.
6.000 V
12.
6.000 V
13.
60.00 V
AC Volts
14.
60.00 V
15.
600.0 V
16.
600.0 V
e
17.
6.000 V
AC Volts + Hz
18.
6.000 V
19.
6.000 V
20.
60.00 V
DC Volts
21.
600.0 V
22.
600.0 V
23.
600.0 mV
m
AC Millivolts
24.
600.0 mV
25.
600.0 mV
DC Millivolts
26.
600.0 mV
27.
 Diode
2.000 V
28.
1000 nF
Capacitance
29.
9999 μF
LoZ
30.
10.00 μF
Capacitance
Applied
Ω
Ω
20
Ω
Ω
250
5 V, 45 Hz
5 V, 1 Hz
50 V, 45 Hz
50 V, 1 Hz
600 V, 45 Hz
600 V, 1 Hz
2 V, 50 kHz
[1]
0 V
5 V
50 V
600 V
-600 V
6 mV, 45 Hz
600 mV, 1 kHz
10 mV
600 mV
1.9 V
Open
900 μF
9 μF
Display Reading
110
114
115
Beeper On
Beeper Off
4.947 to 5.053
4.897 to 5.103
49.47 to 50.53
48.97 to 51.03
593.7 to 606.3
587.7 to 612.3
NA
-0.002 to 0.002
4.973 to 5.027
49.73 to 50.27
596.8 to 603.2
-596.8 to -603.2
5.6 to 6.4
587.7 to 612.3
5.6 to 6.4
587.7 to 612.3
NA
NA
NA
NA
True-rms Multimeter
Performance Tests
[1]
116
117
49.93 to 50.07
1.881 to 1.919
0 to 2
881 to 919
7.18 to 10.82
9

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

113114115116117

Table of Contents