Ohms; Frequency - GE Druck TRX-II User Manual

Multi-function calibrator
Table of Contents

Advertisement

Ohms

Go through the OHM menu and select either the 0 to 400 or the 0 to
2000 Ohms range and enter the DIRECT mode. Resistance is
simulated across the two terminals as shown. Connecting a third or
fourth wire adapts the simulation to 3 or 4 wire input. Adjustment
resolution is respectively 0.01 Ohms and 0.1 Ohms.

Frequency

0 - 12V
= -70mV bias
t =
Ø
Frequencies are symmetrical square waves and -70 mV zero based.
Frequency is set by the
and
keys or numerical key-pad.
Amplitude can be adjusted with the
and
keys between 0 and
24 Volts. Resolution is 0.1 Volt and amplitude is default set to
5.0 Volts. With frequency set at 0 Hz this function becomes available
K319 Issue 2
EN-25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Druck trx-ii is

Table of Contents