Millivolts
Go to millivolts in the DIRECT mode to source any signal between -
10 mV and 100 mV. Adjustments are made with 1 microvolt
resolution. If the adjusted source level can not be maintained the
screen will prompt the warning CHECK LOOP.
Note:'Terminal to test lead' junctions may create e.m.f. microvolts causing
an offset in the actual source signal.
Volts
Go to Volts in the DIRECT mode to source any signal between zero
and 12 Volts. Adjustments are made with 100 microvolts. If the
adjusted source level cannot be maintained the screen prompts the
warning CHECK LOOP.
Milliamps/XMT
warning CHECK LOOP. To source fixed 0, 4, 8, 12, 16 and 20 mA
levels, press
adjust the source current. Press
mode.
EN-24
Use mA terminals to source a current into
a resistor (active mode). Use XMT
terminals to simulate a 2-wire transmitter
(passive mode). Go to milliamps in the
DIRECT mode to source any signal
between 0 and 24 mA. Adjustments are
made with 1 microamps resolution. If the
adjusted source level cannot be
maintained the screen prompts the
key to change to fixed steps. Use
key to return to normal source
and
keys to