Programmable Signal Converter - GE Druck TRX-II User Manual

Multi-function calibrator
Table of Contents

Advertisement

PROGRAMMABLE SIGNAL CONVERTER

With this unique feature any measure signal can be converted into
any on the TRX-II available source signals with full galvanic isolation.
Zero and span settings for both measure and source are fully
programmable.
Setting-up a signal converter
Make all measure and source selections as you would do when you
use measure and source functions separately in the DIRECT source
mode. When the normal working screen is displayed press # and
EXE to open the set-up screen for filling out the measure and source
span. Press EXE again to proceed to the signal converter working
screen. Actual measure and source signals are now displayed.
Note: The signal converter function cannot be used in pulse counting
modes.
I/P Convertor
Make a calibration hook-up as indicated and switch the calibrator on.
Select PRESSURE in the measure menu and mA/XMT in the source
menu. Select the applicable sensor and the desired milliamps source
mode to open the working screen. Read the section "SOURCING OF
ELECTRICAL SIGNALS" on how to adjust the milliamps source.
EN-34
3 - 15 psi
0.2 - 1 bar
Z
S
Supply
pressure
K319 Issue 2