Dräger Medical Narkomed Mobile Technical & Service Manual page 99

Anesthesia system
Hide thumbs Also See for Narkomed Mobile:
Table of Contents

Advertisement

RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM MOBILE
ADJUSTMENT AND CALIBRATION PROCEDURES (continued)
O2 PIPELINE INLET MANIFOLD
SV00325
TEST
GAUGE
ADJUSTING
SCREW
O2 SUPPLY
O2 PIPELINE
PRESSURE ALARM
HOSE
SWITCH
Figure 5-2. Oxygen Supply Pressure Alarm Switch Adjustment
5-5