Choice Of The Appropriate Spacing Of Cuts Acc. To Din En Iso 2409; Choice Of The Appropriate Spacing Of Cuts Acc. To Astm D3359 - Proceq ZCC 2087 Series Instruction Manual

Cross-cut tester
Hide thumbs Also See for ZCC 2087 Series:
Table of Contents

Advertisement

5

Choice of the appropriate spacing of cuts acc. to DIN EN ISO 2409

Film thickness
in μm
0 to 60
0 to 60
above 60 to 120
above 120 to 250
The cross-cut test is not suitable for substrates with a film thickness over 250 μm.
6

Choice of the appropriate spacing of cuts acc. to ASTM D3359

Film thickness
in μm
0 to 50
50 to 125
To test film thickness over 125 μm use test method A, as described in the standard.
Substrate
hard
soft
hard and soft
hard and soft
Film thickness
in mils
0 to 2.0
2.0 to 5
Page 12
Spacing of cuts
in mm
1
2
2
3
Spacing of cuts
in mm
1
2

Advertisement