Configuration Using Tankmaster Winsetup; Introduction; Considerations - Emerson Rosemount 5900 Instruction For Installation, Configuration, And Operation

Proof test function with reference reflector
Table of Contents

Advertisement

Section 3: Configuration of Reference Reflector
November 2019
3.3

Configuration using TankMaster WinSetup

3.3.1

Introduction

The Rosemount 5900 is configured by using the TankMaster Winsetup configuration program.
WinSetup supports standard configuration of the Rosemount 5900 Radar Level Gauge as well as
configuration of the Reference Reflector for Proof Test applications.
See the Tank Gauging
TankMaster WinSetup software to configure a Rosemount Tank Gauging system.
Note
The Proof Test function requires gauge firmware version 1.B9 or higher and Rosemount
TankMaster 6.E1 or higher .
Proof test features
The Rosemount 5900 Proof Test function in TankMaster Winsetup includes the following
functions:
3.3.2

Considerations

The following requirements and recommendations must be considered when using the
Rosemount 5900 Proof Test function:
30
System Configuration Manual
Configure proof test
Perform proof test
View proof test history
Schedule proof tests
Do not perform calibration of Proof Test function during activities in the tank, for
example when it is filled or emptied.
Do not perform calibration of Proof Test function during extreme environmental
conditions.
Proof Test calibration must be repeated whenever configuration of tank geometry
parameters has been changed. This may for example include parameters such as
Calibration Distance or Pipe Diameter.
For Still-Pipes the slots must not be wider than one inch (1")
Manual Supplement
00809-0200-5900, Rev AC
for more information on using the
Configuration of Reference Reflector