Emerson Rosemount 5900 Instruction For Installation, Configuration, And Operation page 18

Proof test function with reference reflector
Table of Contents

Advertisement

Section 2: Installation
November 2019
13.
Figure 2-6. Align the Reference Reflector
14
Ensure that:
the Reference Reflector is correctly aligned towards the center of the tank
inclination of Reference Reflector is less than 2.5°.
Tank wall
Tank center
Parabolic antenna
Reference Reflector (RR)
Manual Supplement
00809-0200-5900, Rev AC
Installation