Maintenance
4.1
Recommended inspection and maintenance
intervals
The following details are manufacturer recommendations. In addition,
comply with the applicable national guidelines.
Type of work to be
performed
Check by the wearer
Visual inspection, function and
leak test
Cleaning and disinfection
Replace the exhalation valve disk
Replace speech diaphragm
1)
Interval may be anhanced to 2 years for air-tight packed facepieces.
2)
Valid for air-tight packed facepieces, other facepieces every 6 months.
106
X
1)
X
X
2)
X
X
X
X
(Use Instructions for NIOSH-approved full facepieces)
4.2
Cleaning and disinfecting
The facepiece must be cleaned after every use so that it is clean
before it is used by a different individual or, if individually assigned, as
often as necessary to keep in a sanitary condition. Facepieces that are
issued to more than one user must be cleaned before being worn by
different individuals.
The facepiece must be disinfected after every use to keep it in a
sanitary condition. It is fully in the user's responsibility to make sure
that the facepiece is totally free of contaminants after this treatment.
Never use a facepiece , that has not been properly disinfected. In case
of doubts dispose of the facepiece in accordance with the local
regulations and use a new one.
WARNING
!
The facepiece body must be thoroughly cleaned so that no
residues from skin care products remain on it.
Otherwise, the elasticity in the sealing range can be lost and
the facepiece will no longer fit correctly.
Non-observance of this warning can cause toxication or
death.
Panorama Nova P