Xantrex XTR 850 WATT SERIES Operating Manual page 50

Xtr 850 watt series programmable dc power supply
Hide thumbs Also See for XTR 850 WATT SERIES:
Table of Contents

Advertisement

Local Operation
3-6
To access the tracking mode for entering voltage and current:
1. Select the VOLTS or AMPS position on the 9-position mode control.
If the set point is blinking, the unit is in coarse tracking mode.
When the VOLTS mode is selected, the voltage set point will
blink in the output voltage display.
When the AMPS mode is selected, the current set point will blink
in the output current display.
2. Use the rotary Adjust/Enter control to adjust the set point.
The adjustments you make to the set point affect the opposing output
value (voltage when in AMPS mode or current when in VOLTS
mode), which will update in the non-blinking display.
3. Press the rotary Adjust/Enter control to enter fine adjust tracking
mode.
The set point blinks faster when the unit is in fine adjust tracking
mode.
4. Use the rotary Adjust/Enter control to fine tune the set point.
5. Once the set point has been selected, press the rotary Adjust/Enter
control to exit tracking mode and return to normal display mode.
To access the select and set entry mode for the voltage and current set
points:
1. Select VOLTS or AMPS position on the 9-position mode control.
2. Press the rotary Adjust/Enter control three times in succession to
enter coarse Pre-Set Adjustment.
3. Press the rotary Adjust/Enter control one more time to enter fine Pre-
Set Adjustment.
4. Press the rotary Adjust/Enter control once again to set the value.
The units can also timeout (adjustable) to exit the menu. Once Pre-Set
Mode is entered crpc and fnpc will appear in the voltage display.
M370046-01

Advertisement

Table of Contents
loading

Table of Contents