Query for Analog Current Input Level
To query for analog current input level from non-isolated input:
1. Turn the 9-position mode control to the CAP position or press the
2. If the mode is currently nonE then select any other APG mode. If this
3. Press the rotary Adjust/Enter control to select the APG mode.
4. Press the rotary Adjust/Enter control to return to normal mode.
5. If the APG mode was modified in step 2, execute the procedure to
The SCPI command (s) for these instructions are:
Quick Tip
Remote operation
[:]SYSTem[<channel>]:REMote:SOURce:CURRent <?|LOCal
uses these SCPI
|AVOLtage|IAVoltage|ARESistive|IAResistive>
commands. See
page 5–1.
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel:CURRent
[:ISOlated]
[:]MEASure[<channel>][:SCALar]:APRogram:CURRent[:DC]?
M370046-01
rotary Adjust/Enter control if the mode control is already at the CAP
position.
CAPr is displayed on the output voltage display.
step is done, be sure to restore the APG setting to nonE when the
operation is complete. Otherwise, do not rotate the rotary Adjust/
Enter control to keep the current APG mode.
The output voltage display should read LE C and the output current
display will display APG level.
deactivate the APG voltage mode.
Analog Programming Mode
4-13
4