Remote Operation
Running an Auto Sequence Program:
5-64
4. From the Hyper Terminal (or any other terminal program) select the
Transfer>Send Text File...
5. Navigate to and select the text file that was previously stored. Click
the Open button.
6. Execute the program recording stop command:
:PROG:STOP
7. Execute the readback command to verify the entire program was
transferred correctly:
PROG:READ?
8. Set the data rate back to the original speed.
9. Close the connection to the HyperTerminal program and configure
the connection data rate to be the original speed.
Once an auto sequence program has been stored, there are a few
commands that can be used to run the program. Once a program begins
running, it will continue running until completion and then repeat the
number of times specified by the repeat command. The default at power
on is to have repeat set to 1.
Commands:
*TRG
[:]INITiate[<channel>][:IMMediate]
[:]PROGram[<channel>]:STATe {?|RUN|STOP|PAUSe}
Where:
RUN causes the program to begin execution.
STOP causes the program to stop execution
PAUSe causes the program to stop executing at the current command. If a
subsequent run is sent, the program will resume executing with the next
command in the program.
For Example:
:PROG:STAR
*IDN?
*IDN?
:PROG:STOP
M370046-01