Download Print this page

Hach NA5600 sc Na+ Operation Manual page 27

Hide thumbs Also See for NA5600 sc Na+:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Option
NUMBER OF HOURS
NUMBER OF READINGS
6.4 Save data or settings to an SD card
Save the data logs to an SD card to use the data on a PC as necessary. Save the analyzer settings
to an SD card so the settings can be restored later or copied to another analyzer as necessary.
Items to collect:
• SD card (2 GB or more)
• PC with SD card slot
1. Put an SD card in the SD card slot (2 GB minimum). Refer to
2. Push menu, then select SD CARD SETUP.
3. Select an option.
Option
UPGRADE
SOFTWARE
SAVE LOGS
MANAGE
CONFIGURATION
WORK WITH
DEVICES
4. When done, remove the SD card from the analyzer.
5. Install the cover for the SD card slot to keep the environmental rating of the enclosure.
Description
Shows the data recorded within the selected number of hours before now.
Shows the selected number of data points.
Description
Note: The UPGRADE SOFTWARE option only shows when a software update file is
on the SD card.
Installs the software update file that is on the SD card. Refer to
software version
on page 28.
Saves the data log file to the HACH/Logs/ANALYZER_xxxx folder on the SD card.
Open the data log file, ANALYZER_NAME_DL.xml, with Internet Explorer or Excel.
Saves the event log file to the HACH/Logs/ANALYZER_xxxx folder on the SD card in
CSV (comma-separated value) file format. Open the event log file,
ANALYZER_NAME_EL.csv, with Excel.
Options: LAST DAY, LAST WEEK, LAST MONTH, ALL or NEW.
Note: To save the other logs files to the SD card, refer to the WORK WITH DEVICES
option.
BACKUP SETTINGS—Saves the analyzer settings to the SD card. TRANSFER
SETTINGS—Installs the analyzer setting saved to the SD card on the analyzer.
READ DEVICE FILES—Saves the selected device data to the HACH/Devices folder
on the SD card in CSV file format. Options: GRAB SAMPLE DATA, CAL HISTORY,
SENSOR DIAG, MEASUREMENT DATA (curve data for calibrations and grab
sample measurements) and SERVICE HISTORY. WRITE DEVICE FILE—Installs a
new version of the measurement cycle script.
Note: The WRITE DEVICE FILE option only shows when a new version of the
measurement cycle script is on the SD card.
Figure
11.
Install the latest
English 27

Advertisement

loading