Download Print this page

Hach NA5600 sc Na+ Operation Manual page 20

Hide thumbs Also See for NA5600 sc Na+:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
• SCHEDULER function (refer to
Option
HOLD OUTPUTS Holds or transfers outputs for the selected channels.
RUN DAYS
START TIME
INTERVAL
DURATION
OFF DELAY
Figure 10 Scheduler function
1 Duration
2 OFF delay
• WARNING function
Option
WARNING LEVEL Sets the relay to on when the selected warning(s) occurs. Push the LEFT arrow to select
• PROCESS EVENT function
Option
Description
SELECT
Sets the relay to on when the selected process event(s) occurs. Push the LEFT arrow to select
EVENTS
or unselect a checkbox.
MEASURING 1, 2, 3 or 4—Sets the relay to on during the measurement cycle of Channel 1, 2,
3 or 4.
CALIBRATE—Sets the relay to on during calibration.
SHUTDOWN—Sets the relay to on when in shutdown mode.
STARTUP—Sets the relay to on during the startup cycle.
GRAB SAMPLE—Sets the relay to on during grab sample measurement.
MARK END OF MEASURE—Sets the relay to on for 1 second at the end of each
measurement cycle.
20 English
Figure
Description
Sets the days that the relay operates.
Sets the start time.
Sets the time between activation cycles (0 to 999 seconds, default: 0).
Sets the period of time the relay is energized (0 to 999 seconds, default: 0).
Sets the time for additional hold/output time after the relay has been turned off (0 to
999 seconds, default: 0).
Description
or unselect a checkbox.
10)
3 Interval
4 Time (x-axis)

Advertisement

loading